Double Beam Spectrophotometer 6850 supplied with single 10x10 mm cuvette holder, power cable, PC Software on CD ROM_1519144
Representative image only

Double Beam Spectrophotometer 6850 supplied with single 10x10 mm cuvette holder, power cable, PC Software on CD ROM

Spectrophotometer Model 6850
Double beam spectrophotometer with a variable spectral bandwidth. The highly stable optics and two detectors measure the sample and reference simultaneously optimising measurement accuracy. The 6850 has measurement modes for photometrics, concentration, multi-wavelength, spectrum scanning, kinetics, quantitation, DNA/RNA and protein analysis. Jenway Prism PC software is supplied as standard and offers additional functionality with preloaded methods for DNA/RNA and protein analysis, as well as extensive post-measurement tools, unlimited results saving and easy export of data. The 6850 double beam spectrophotometer is ideal for quality control, general research, pharmaceutical, biochemical and clinical laboratory applications. Installation Qualification and Operation Qualification (IQ/OQ) is available for the 6850.

-Double beam spectrophotometer with highly stable optics
-Variable spectral bandwidth 0.5, 1, 2, 4, 5 nm
-Analysis: Auto peaks and valleys, zoom, addition, subtraction, peak ratios, smoothing, area under curve, wavelength table, derivatives, overlay with PC software
-Integrated user interface
-USB port for data storage
-Conforms to European Pharmacopeia requirements
-Jenway Prism PC software included as standard
-1 year warranty
Specifications
Optical system:
Double beam
Wavelength range:
Resolution:
Accuracy: 190 to 1100 nm
0.1 nm
± 0.3 nm (at 0.5 and 1 nm bandwidth)
± 0.5 nm (at 2, 4 and 5 nm bandwidth)
Reproducibility: ± 0.2 nm
Photometric range: -0.3 to 3.0 A
0 to 200 %T
Accuracy: ± 0.002 A (0 to 0.5 A)
± 0.3 % T (0 to 100 %T)
Reproducibility: ±0.001 Abs (0 to 0.5 Abs)
±0.002 Abs (0.5 to 1.0 Abs)
Resolution:
Stray light: 0.1 %T, 0.001 A
< 0.05 %T at 220 and 360 nm
Concentration range:
Quantitation points: 0 to 99999
Up to 3 wavelengths
Kinetics: Up to 12 h with time intervals of 0.1, 0.2, 0.5, 1, 2, 5, 10 or 30 s
Spectrum:
Scan Speed:
Scan interval: Any range between 190 and 1100 nm
100 to 2000 nm/min
0.1, 0.2, 0.5, 1, 2 or 5 nm
Light source: Tungsten and Deuterium lamp
Outputs: USB and parallel
Dimensions (W x D x H): 600 x 450 x 200 mm
Weight: 22 kg
Supply requirements: 230 V, 50/60 Hz

Over 40,000 Customers Trust John Morris