Active microprocessor control for humidifying and dehumidifying (20 – 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter. Active humidity control creates a controlled environment for accelerated life tests and 85/85 tests.
Working-temperature range: +7 above ambient temperature up to +90 °C
Setting temperature range: +18 to +90 °C
Setting accuracy temperature: 0.1 °C
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error